|
FHWA-SA-95-038
FWD CALIBRATION CENTERS
These guidelines are provided to ensure that a trip to a Strategic Highway Research Program (SHRP) Falling Weight Deflectometer (FWD) calibration center is as productive and problem-free as possible. Your cooperation in following the guidelines is greatly appreciated by calibration center staff. If you have any questions about FWD calibration, contact a Regional Calibration Center Operator or the Federal Highway Administration (FHWA) Long-Term Pavement Performance (LTPP) Division Office (see Calibration Center Contacts). A copy of the complete calibration protocol can be obtained from the LTPP Division Office. The SHRP FWD Calibration Centers were developed to calibrate SHRP's Dynatest FWDs. The procedures and equipment have also been used to calibrate KUAB FWDs, and additional adaptations to provide calibration of Phonix and JILS FWDs are under development. Should you wish to calibrate a Phonix or JILS FWD, please contact Mr. Eric Weaver at the FHWA LTPP Division Office. WHAT TO EXPECT. . . From the Calibration Process The calibration process yields a set of multipliers (one for each sensor calibrated) that are used in the FWD software to adjust raw sensor readings (see figure 1). If you think of deflection measurements as shots at a target, the data from the uncalibrated sensors of an FWD could be represented by figure la. They may or may not fall within the specified accuracy of 2 percent, represented by the outermost ring on the target, and they may or may not be systematically biased (all to one side of the bull's-eye). After reference calibration (see figures 2 and 3), the readings will be closer to the bull's-eye and unbiased (randomly scattered around the bull's-eye), as shown in figure lb. Finally, after relative calibration, the readings will be randomly scattered and even closer to the true deflection (represented by the center of the bull's-eye), as shown in figure lc. Figure 2 illustrates the reference calibration of the load measurement system. Figure 3 illustrates the reference calibration of an FWD deflection sensor.
From the Calibration Center Staff The calibration center operator runs the reference calibration equipment and software. He or she will ask FWD operators to provide printouts from the FWD software to document the starting condition and serial numbers of the FWD sensors, and will then tell FWD operators what to do for each step of the calibration. The actual operation of the FWD during the calibration process is the responsibility of the FWD operator. When the calibration process is complete, the calibration center operator returns the calibration data diskette1 to the FWD operator and provides the operator with a printed calibration report. Distribution of the report is the FWD operator's responsibility. One copy of the report should be transmitted to the FHWA LTPP Division Office. 1 Refer to Item 6 of General FWD Equipment Requirement section later in this document.
![]()
WHAT NOT TO EXPECT...
From the Calibration Results
Do not expect perfect calibration results. Generally, gain factors for an FWD are between 0.980 and 1.020. This range corresponds to the specified accuracy of most FWD sensors. Factors outside this range may indicate a damaged sensor or other equipment problems. If you doubt calibration results for one or more sensors, and if time permits, repeat the calibration to verify the results. In general, results from two successive calibrations of a given sensor should agree to within ± 0.003.
From the Calibration Center
Do not expect the calibration center to provide spare parts or FWD repairs. Host States operate the FWD calibration centers on a voluntary basis, and the calibration center operators have other duties.
![]()
SPECIFIC FWD EQUIPMENT REQUIREMENTS FOR CALIBRATION Before going to the calibration center, you should be prepared with the following items:
3 seating drops at
height 3 followed by a pause, Dynatest FWDs require Version 10.0 or higher of the Dynatest field program. KUAB FWDs require operational program SFWD Version 4.0 or higher. For other FWDs, please consult your users guide, or check with the manufacturer. Figures 4 and 5 present sample reference and relative calibration test setups for LTPP Dynatest equipment.
2 It is desirable, but not essential, to program six drops at each drop height rather than five. This provides a margin for error in coordination between the FWD operator and calibration system operator.
GENERAL FWD EQUIPMENT REQUIREMENTS
Note: Complete the
Pre-trip
Checklist
CALIBRATION CENTER PROCEDURES
3 If a load of 6000 lbs (± 10%) at drop height 1 cannot be achieved with the same FWD weight packages used for the remaining target loads, use lowest practical drop height.
|
|||||
|
FEDERAL HIGHWAY ADMINISTRATION |
|||||
|
Attention:
Mr. Eric Weaver
Turner-Fairbank Highway Research Center 6300 Georgetown Pike, HRDI-13 McLean, VA 22101-2296 Phone: 202-493-3159 Fax: 202-493-3161 |
TFHRC Home | FHWA Home | Feedback United States Department of Transportation - Federal Highway Administration |